Microincinerators ST-14A

Features:

  • Infrared heat delivered within the ceramic tube
  • No flames, gasses, or splatters
  • Inoculating loops & needles sterile in 3-5 seconds
  • Ceramic heating tube quickly reaches 900°C
  • Lifetime heating element and stable base
  • (ST-14A only) The angle of the heating tube is adjustable to the user’s preference

 

 

 

 

The microincinerator is designed to sterilize metal inoculating loops and needles without using an open flame. By using infrared heat inside a ceramic tube, the product protects your laboratory technicians from dangerous gases, flames, and splatters. Once it reaches the optimum sterilization temperature (900℃), the loop or needle is sterilized within 3-5 seconds.

Specifications:

Model ST-14A
Picture
Max. Temp in Operation Mode 900℃± 50℃
Max. Temp in Standby Mode 650℃± 50℃
Max. Sample Diameter Φ14mm (0.55″)
Depth of Heated Area 130mm (5.11″)
Sterilization Time 3~5s
Angle Adjustment Yes
Ambient Temperature 4℃~60℃
Relative Humidity ≤ 90% Non-condensing
Power 200W
Power Supply AC110V/ 60Hz
Weight 1.2kg (2.64lbs)

 

Accessories

Name Part Number (Previous Part Number) Picture
Inoculation Loop Holder STA01(S1121-1)  

 

Download our Product Datasheet (PDF):

Microincinerator-Datasheet-7-15-21-compressed


Request Quote from Distributor or more information:
info@crystalindustries.com